Artison LRS User Manual Page 4

  • Download
  • Add to my manuals
  • Print
  • Page
    / 10
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 3
The TR8818A digitizer, and 6103 amplifier module are all fully programmable for automated testing,
remote operation, and quick setups. Remote commands set the digitizers' sample clock rate, pre-trigger
memory fraction, waveform record size, and input offset. Remote commands can also arm and trigger
the digitizers. A Look-At-Me flag gets set when data is ready for readout. Look-At-Me (LAM) gets
translated to a GPIB Service Request (SRQ) by the LeCroy 8901A.
SPECIFICATIONS
ANALOG INPUT
Channels: One.
Connector: BNC type coaxial.
Impedance: 50 ohm ±3% DC; < 7% reflection from 3 nsec edge.
Bandwidth: DC to > 120 MHz guaranteed.
DC Offset Voltage: Programmable over ±1/2 full scale in 256 steps; ±0.1% full scale/°C.
Coupling: DC.
Overshoot: < 5% typical for step input with or without 6103 amplifier.
Overload Recovery: ±1% F.S. in 25 nsec from a 1.5X overdrive pulse of < 1 µsec duration.
Overload Protection: ±2.5 VDC INPUT, ±7.5 VDC HIGH LEVEL INPUT; ±500 Vpk for 30 nsec;
±100 Vpk for 100 µsec with 50 ohm source.
ANALOG-TO-DIGITAL CONVERSION
Input Amplitude Range: 510 mV p-p (2 mV/LSB) ±1% F.S. ±0.1%/°C.
Single Shot Resolution: 8 bit (offset binary code).
DC Accuracy: Errors < ±1.5 LSB + 0.5 LSB quantiza tion error from best linear fit over entire range.
AC Accuracy: A reconstructed digitized pure sine wave has a typical dynamic accuracy as follows:
Si
g
nal Frequenc
y
Si
g
nal
Amplitude
Effective Bits
Si
g
nal-to-Noise Ratio
(SNR)*
DC5 MHz 80% F.S. 6.5 47 dB
520 MHz 80% F.S. 6 45 dB
Page 3 of 8tr8818.htm
12/15/2008htt
p
://www.lecro
y
.com/lrs/dsheets/8818.htm
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Page view 3
1 2 3 4 5 6 7 8 9 10

Comments to this Manuals